The Biotest RCS Compressed Gas Adapter (Article No. 940 720), in conjunction with the RCS Plus, RCS High Flow or RCS Isolator Microbial Airborne Sampler, is used to investigate the microbiological content of pressurized gasses according to ISO 14698. It is routinely used in areas where, due to legal requirements or individual standards, specified air micobial counts must not be exceeded.
Technical Data:
Material:
- Costructed primarily of anodized aluminum, the CG Adapter is
easily disinfected by commercially availabe detergents and/or
by autoclaving.
- In addition the stainless steel connection parts can be
sterilized with H2O2.
Range of Application:
- 1 bar +/- 10% (other pressures in conjunction with Nozzle Set)
- The adapter is suitable only for non-flammable and non-toxic gasses.
Flow Rate:
- At a pressure of 1 bar the flow rate is approximately 180 lpm
Sample Volume:
- Under ISO 14698 recommendations 1 m³ of air should be sampled.
Additional Accessories:
- Nozzle Set for RCS Compressed Gas Adapter (Article No. 940 725) for adjustment to a pressure range of 0.1-7 bar